Incorrect GE Hip Analysis

In this left hip scan, the scan did not go down far enough into the femoral shaft, therefore the total hip region of interest is incorrect.

Case Description:

In this GE Healthcare scan of the left hip, the scan did not go down far enough in the femoral shaft. The region of interest, for the total hip, is therefore incorrectly placed and no total hip data is available.  Notice that the triangle for the total hip region of interest is flipped upwards.

Credit:

Sarah L Morgan, MD, RD, CCD, The University of Alabama at Birmingham

References:

• Watts, N.B., Fundamentals and pitfalls of bone densitometry using dual-energy X-ray absorptiometry (DXA). Osteoporos Int, 2004. 15(11): p. 847-54.
• Bojinca, V., D. Opris, and M. Bonjinca, Artifacts and pitfalls in DXA scan images and interpretation. J Clin Densitom, 2012. 15(4): p. 486-487.
• Kiraç, F.S., D. Yüksel, and O.T. Yaylali, Pitfalls in the measurement of bone mineral density by the dual-energy X-ray absorptiometric method. Clin Nucl Med, 2001. 26(10): p. 874-5.
• Hansen, K., et al., DXA Errors are Common and Likely Adversely Affect Clinical Care: DXA Quality Improvement is Needed. J Bone Miner Res 2016. 31((Suppl 1) Available at http://www.asbmr.org/ItineraryBuilder/Presentation Detail.aspx?pid=83c01c31-237b-4f07-81a5-1eeb2a7968aa&ptag=AuthorDetail&aid=00000000-0000-0000-0000-000000000000. ).
• Bendavid, E.J., et al., Frequency and magnitude of analysis erros of bone density results assessed by dual x-ray absorptiometry. J Bone Miner Res, 1993. 8 (Suppl 1): p. S345.
• Binkley, N., et al., Error prevalence in DXA performance and reporting: Improving DXA quality is essential. . J Clin Densitom, 2016. ? (? ): p.? .
• El Maghraoui, A. and C. Roux, DXA scanning in clinical practice. QJM, 2008. 101(8): p. 605-17.
• Fenton, J.J., et al., Osteoporosis Overtreatment in a Regional Health Care System. JAMA Intern Med, 2016. 176(3): p. 391-3.
• Garg, M.K. and S. Kharb, Dual energy X-ray absorptiometry: Pitfalls in measurement and interpretation of bone mineral density. Indian J Endocrinol Metab, 2013. 17(2): p. 203-10.
• Kaleta, M. and S. Wronski, The most common errors in the densitometric diagnosis of osteoporosis. Ortop Traumatol Rehabil, 2001. 3(3): p. 338-44.
• Kim, T.Y. and A.L. Schafer, Variability in DXA Reporting and Other Challenges in Osteoporosis Evaluation. JAMA Intern Med, 2016. 176(3): p. 393-5.
• Krueger, D., et al., DXA Errors Are Common and Reduced by Use of a Reporting Template. J Clin Densitom, 2019. 22(1): p. 115-124.
• Lewiecki, E.M., N. Binkley, and S.M. Petak, DXA quality matters. J Clin Densitom, 2006. 9(4): p. 388-92.
• Lewiecki, E.M. and N.E. Lane, Common mistakes in the clinical use of bone mineral density testing. Nat Clin Pract Rheumatol, 2008. 4(12): p. 667-74.
• Maldonado, G., et al., Common errors in dual-energy X-ray absorptiometry scans in imaging centers in Ecuador. Arch Osteoporos, 2020. 15(1): p. 6.
• Martineau, P., M. S.L., and W.D. Leslie, Bone mineral densitometry rerporting: Pearls and pitfalls. . Can Assoc Radiol J 2020.
• Messina, C., et al., Prevalence and type of errors in dual-energy x-ray absorptiometry. Eur Radiol, 2015. 25(5): p. 1504-11.
• Lewiecki, E.M., et al., Best Practices for Dual-Energy X-ray Absorptiometry Measurement and Reporting: International Society for Clinical Densitometry Guidance. J Clin Densitom, 2016. 19(2): p. 127-40.
• El-Hajj Fuleihan, G., et al., A national random survey of bone mineral density reporting in the United States. J Clin Densitom, 2002. 5(1): p. 3-9.
• Johnston, R., et al., Quality assessment of bone density testing by DXA – Evaluation of technical and reporting deficiencies identified at a tertiary osteoporosis clinic. . J Clin Densitom, 2016. 19: p. 538-9.
• England, J.R. and P.M. Colletti, Automated Reporting of DXA Studies Using a Custom-Built Computer Program. Clin Nucl Med, 2018. 43(6): p. 474-475.
• Wachsmann, J., et al., Electronic Medical Record Integration for Streamlined DXA Reporting. J Digit Imaging, 2018. 31(2): p. 159-166.